TUTORIAL 4671

Improve Sensor Performance and SNR in Pulse Oximeter Designs

By:  John DiCristina

Abstract: This tutorial discusses the different types of pulse oximeters including optical and reflective techniques and the design considerations taken when building high-end portable bedside monitors or mid and low-end battery-powered models. The discussion covers light absorption, sensor calibration, and details specifics within the transmit path leading to in-depth discovery of photodiode, interface, signal conversion, and data conversion characteristics. A functional block diagram and part recommendations are also included.

Overview

Pulse oximeters noninvasively measure or continuously monitor oxygen saturation in arterial blood to ensure that there is sufficient oxygenation. Typical applications include people with respiratory problems who are under anesthesia, neonates, and critically ill patients.
In a pulse oximetry system, a clip with optical electronics is usually attached to a finger, toe, or ear so that light can be transmitted through the skin from one side of the clip and received on the other side with a photodiode. Good arterial blood flow is required to measure oxygen saturation. Most applications use transmissive optical techniques, but there are some applications that use reflective techniques.

Oximeter Types

Available in different shapes and sizes, pulse oximeters fall within three main categories.
At the high end are portable bedside patient monitors, which have a removable cable that includes a clip containing the optics used to measure oxygen saturation. These models are either dedicated to pulse oximetry or they monitor multiple vital parameters such as pulse, blood pressure, respiratory rate, and temperature. These high-end models demand high performance, with a low-noise signal path being the most important design parameter. Power is a secondary consideration, since the monitor is line powered most of the time and only runs off of battery power when the patient is mobile. Size is not a driving factor in these designs.
Mid-range handheld pulse oximeter.
Mid-range handheld pulse oximeter.
Mid-range models are typically battery-powered handheld units with a removable cable. Performance is still important, but it needs to be balanced with size and power constraints. While handheld units run off of batteries all of the time, they have the space to include decently sized batteries.
Low-end fingertip pulse oximeter.
Low-end fingertip pulse oximeter.
At the low end are fingertip models, which combine the base unit and cable into a clip that is large enough to cover a finger and house the electronics and batteries. The small size means that performance may need to be sacrificed to preserve the small form factor and achieve the desired battery life.
Functional block diagram of a pulse oximeter.
Functional block diagram of a pulse oximeter. For a list of Maxim's recommended solutions for pulse oximeters, please visit: www.maximintegrated.com/pulseox.

Principle of Operation

Operation is based on the light-absorption characteristics of hemoglobin in the blood. Oxygenated hemoglobin absorbs more infrared light than red light, and deoxygenated hemoglobin absorbs more red light than infrared light. Thus, red and infrared LEDs in the oximeter alternately transmit light, and a photodiode receives the light that is not absorbed. The ratio of the red and infrared light received by the photodiode is used to calculate the percentage of oxygen in the blood. Based on the pulsatile nature of arterial blood flow, the pulse rate and strength are also determined and displayed during the measurement cycle.

Sensor Cable Calibration

Pulse oximeter sensor cables are removable and often disposable. The performance of the LEDs and the photodiode in the cable can vary due to either manufacturing tolerances or alternate versions of the sensor. The performance of the pulse oximeter can be improved by incorporating nonvolatile memory such as EPROM or EEPROM into the sensor cable to store calibration coefficients. This calibration data is used by the base unit to optimize the performance of the particular optical components in the sensor.
Because the electromechanical connector for the sensor cable typically has limited pins, system designers must add this calibration functionality with minimal interconnect. The simplest and most robust method to do this is to add a 1-Wire memory device to the cable connector. The 1-Wire protocol is a proven serial interface that uses only one connector pin to add calibration and other functionality to pin-limited designs.
Additionally, since each 1-Wire product is designed with high ESD protection built in, designers do not need to use external protection diodes.
An added benefit of 1-Wire memory devices is that they can be used to pass updates or new algorithms to the base unit, thereby providing a vehicle for in-field upgrades. They can also enable aftermarket control by allowing only approved sensor cables to be used. A unique 64-bit serial identification number is inscribed in each 1-Wire device. By writing to a lockout bit in the memory device, manufacturers are able to control how many times the sensor can be used before replacement is required. Replacement is driven by safety concerns and, more specifically, by patient contamination issues.

Pulse Oximeter Solutions

Transmit Path: Driving the LEDs

The red and infrared LEDs are driven alternately with a precise current; a quiet time occupies the interval between each alternating pulse. The repetition rate is not fast, usually well under 10kHz. The duty cycle is also low to keep overall power down and to allow ambient-light measurement while both LEDs are off. Most of the time the LEDs are at the end of a cable, so system designers prefer to drive the LEDs with current from the base unit. This approach reduces current errors and requires fewer lines in the cable, which, in turn, reduces cost and weight. The current used to drive the LEDs must have minimal noise because any noise in the bandwidth of interest will degrade the signal-to-noise ratio (SNR) of the system and decrease its ability to measure low oxygen-saturation levels in critically ill patients.
Off-the-shelf LED drivers are not typically used in this application for several reasons. It is difficult to accurately and precisely program the LED current with the required resolution, since the backlighting applications for which they are designed do not require high resolution. Also, if a switching topology is used for the LED driver, then the current noise will be too high. It is possible that an LED driver with a linear topology could be used, but this approach is not ideal due to the lack of resolution in programming the current.
The LEDs can be driven independently with a line from the base unit for each LED along with a shared ground. Alternatively, to save connector pins and lines, the LEDs can be driven in a back-to-back or antiparallel configuration. Figure 1 shows a method for driving the two LEDs in the cable using two digital-to-analog converters (DACs) to independently set the currents in the LEDs. The advantage of using two DACs is that they can be heavily filtered to reduce the noise because their output is relatively static. LED modulation is performed after the filtering of the DACs in this topology.
Figure 1. Simplified red and infrared LED bias and control circuit.
Figure 1. Simplified red and infrared LED bias and control circuit.
In all topologies, the operational amplifier used in the force-sense (FS) configuration needs to have low voltage noise and high enough bandwidth to modulate the LEDs. A single DAC can be used to reduce cost, but the performance may suffer if the noise increases. The peak current through each LED varies according to the characteristics of each LED and photodiode, and with the duration of each pulse. In practice, the peak current ranges from tens of mA to hundreds of mA across all oximeters, though individual meters will use a narrower range of currents.

Receive Path: Photodiode Interface, Signal Conditioning, and Data Conversion

A photodiode receives both ambient and modulated light from the red and infrared LEDs, and then generates a current that will be measured over time to determine oxygen saturation. The photodiode current is most often converted to a voltage using an operational amplifier in a transimpedance configuration, also called a transimpedance amplifier (TIA). The light received by the photodiode largely comprises ambient light that is not useful for determining oxygen saturation or pulse. The small amount of red or infrared light not absorbed by the tissue, venous blood, or arterial blood is the signal of interest and is buried in the ambient light.
Figure 2 presents a simplified circuit for converting the photodiode current to a voltage, for signal conditioning, and for analog-to-digital conversion. This circuit converts both the red and infrared signals together and then processes them in the digital domain. Alternate topologies exist where the red and infrared light are separated in the analog domain using synchronous demodulation. This technique allows the use of slow but high-resolution analog-to-digital converters (ADCs); however, it is poor at detecting motion artifacts.
Figure 2. Simplified photodiode receive path circuit.
Figure 2. Simplified photodiode receive path circuit.
The key specifications for the TIA are extremely low input current, input current noise, and input voltage noise, as well as high-voltage operation. These characteristics are necessary to maximize the SNR so that the small currents of interest can be measured amid the large ambient currents. High-voltage operation means that a larger feedback resistor can be used to easily amplify the ambient and received LED current before removing the ambient portion with a highpass filter. The remaining small signal of interest is then amplified to maximize the ADC's dynamic range. This gain stage should be programmable to compensate for changing environmental factors and the aging of optical components.
The key specifications for the ADC are high SNR and sample rate. The sample rate should be fast enough to capture the modulated signal and perform the required digital processing, which is different for each manufacturer.

Display and Backlighting

If a bedside pulse oximeter is part of a multiparameter patient monitor, it does not require its own display because it uses the large color display incorporated in the patient monitor. Bedside monitors dedicated to pulse oximetry will usually have a medium-sized color display, or they can incorporate a touch-sensitive display to enhance the user interface and reduce the number of dedicated buttons. Handheld models have either a simple LED or LCD segment display or a more advanced dotmatrix LCD display. Fingertip units have the space and power only for a small and simple LED or LCD display. Simple displays can be driven by the display drivers found in many microcontrollers; dot-matrix or more advanced displays will require dedicated circuitry or a turnkey display solution. Backlighting is not required for LED displays; a few white LEDs can be used to backlight small LCD displays.

Data Interface

Most handheld and bedside models have the ability to interface with a computer. This data transfer is usually performed by a health-care provider or technician and not by the patient. Traditionally, the data interface has used RS-232, and the format has been proprietary to the manufacturer. Pulse oximeters are now migrating toward USB or wireless (Bluetooth®, Wi-Fi®) interfaces, and in the future oximeters will be able to securely transfer data to any device with these standard interfaces.

Audible Alarms

Audible alarms range from simple buzzers to speakers outputting multiple tones, levels, and patterns. A simple buzzer can be driven by one or two microcontroller port pins with pulse-width modulation (PWM) capability. More advanced audible alarms can be achieved by adding an audio DAC along with a speaker amplifier.

Power and Battery Management

A pulse oximeter needs to generate multiple power-supply rails. Separate analog (3V) and often multiple digital (3V, 1.8V) supplies are required if there are low-voltage cores requiring their own supply. A separate clean supply for powering the LEDs (5V) is usually required due to the power, headroom, and noise considerations. A high voltage (+30V or -30V) is sometimes required for the TIA connected to the photodiode to increase the SNR and, in some cases, for the display.
Fingertip models use two primary alkaline batteries in the AAA form factor while handheld models use either two primary AA alkaline or rechargeable batteries. Both models require a combination of step-up switching regulators, low-dropout linear regulators (LDOs), and sometimes an inverting switching regulator. Bedside monitors are mainly line powered but will use lithium-ion (Li+) or nickel-metal-hydride (NiMH) rechargeable backup batteries. A regulated 5V DC voltage is usually supplied, so the power-supply design is different than the fingertip and handheld models. Battery charging and fuel gauges are needed for the models with rechargeable batteries, as well as authentication for safety and after-market control if the batteries are removable.

Electrostatic Discharge

Electrostatic discharge (ESD) is a concern where the sensor cable plugs in. This potential problem can be addressed with dedicated ESD line protectors, as well as careful design and layout of the board. The use of ESD-protected interfaces allows designers to avoid the space and cost associated with discrete protection components. ESD is also possible at the data port and printer port, if the oximeter is equipped with one. Air discharge may also need to be addressed for the buttons and display through small openings in the casing, but this depends on the individual oximeter design.

1-Wire is a registered trademark of Maxim Integrated Products, Inc.

The Bluetooth word mark and logos are registered trademarks owned by Bluetooth SIG, Inc. and any use of such marks by Maxim is under license.

Wi-Fi is a registered certification mark of Wi-Fi Alliance Corporation.



Related Parts
DS2431 1024-Bit 1-Wire EEPROM Free Samples  
DS2460 SHA-1 Coprocessor with EEPROM Free Samples  
DS2480B Serial to 1-Wire Line Driver Free Samples  
DS2482-100 Single-Channel 1-Wire Master Free Samples  
DS2502 1Kb Add-Only Memory Free Samples  
DS2756 High-Accuracy Battery Fuel Gauge with Programmable Suspend Mode Free Samples  
DS2780 Stand-Alone Fuel Gauge IC Free Samples  
DS2784 1-Cell Fuel Gauge with FuelPack, Protector, and SHA-1 Authentication Free Samples  
DS2784 1-Cell Fuel Gauge with FuelPack, Protector, and SHA-1 Authentication Free Samples  
DS28E01-100 1Kb Protected 1-Wire EEPROM with SHA-1 Engine Free Samples  
DS28E02 1-Wire SHA-1 Authenticated 1Kb EEPROM with 1.8V Operation Free Samples  
DS28E05 1-Wire EEPROM Free Samples  
MAX1162 16-Bit, +5V, 200ksps ADC with 10µA Shutdown Free Samples  
MAX1167 Multichannel, 16-Bit, 200ksps Analog-to-Digital Converters Free Samples  
MAX1168 Multichannel, 16-Bit, 200ksps Analog-to-Digital Converters Free Samples  
MAX11800 Low-Power, Ultra-Small Resistive Touch-Screen Controllers with I²C/SPI Interface  
MAX1393 1.5V to 3.6V, 312.5ksps, 1-Channel True-Differential/2-Channel Single-Ended, 12-Bit, SAR ADCs  
MAX1396 1.5V to 3.6V, 312.5ksps, 1-Channel True-Differential/2-Channel Single-Ended, 12-Bit, SAR ADCs  
MAX15029 1.425V to 3.6V Input, 500mA Low-Dropout Regulators with BIAS Input Free Samples  
MAX16033 Low-Power Battery-Backup Circuits in Small µDFN Packages Free Samples  
MAX16054 On/Off Controller with Debounce and ±15kV ESD Protection Free Samples  
MAX16814 Integrated, 4-Channel, High-Brightness LED Driver with High-Voltage DC-DC Controller Free Samples  
MAX16826 Programmable, Four-String HB LED Driver with Output-Voltage Optimization and Fault Detection Free Samples  
MAX16838 Integrated, 2-Channel, High-Brightness LED Driver with High-Voltage Boost and SEPIC Controller Free Samples  
MAX17043 1-Cell/2-Cell Fuel Gauge with ModelGauge and Low-Battery Alert  
MAX1916 Low-Dropout, Constant-Current Triple White LED Bias Supply Free Samples  
MAX2829 Single-/Dual-Band 802.11a/b/g World-Band Transceiver ICs Free Samples  
MAX2830 2.4GHz to 2.5GHz 802.11g/b RF Transceiver with PA and Rx/Tx/Diversity Switch Free Samples  
MAX2831 2.4GHz to 2.5GHz, 802.11g RF Transceivers with Integrated PA  
MAX3207E Dual, Quad, and Hex High-Speed Differential ESD-Protection ICs Free Samples  
MAX3208E Dual, Quad, and Hex High-Speed Differential ESD-Protection ICs Free Samples  
MAX4252 UCSP, Single-Supply, Low-Noise, Low-Distortion, Rail-to-Rail Op Amps Free Samples  
MAX5137 Pin-/Software-Compatible, 16-/12-Bit, Voltage-Output DACs Free Samples  
MAX5535 Dual, Ultra-Low-Power, 12-Bit, Voltage-Output DACs Free Samples  
MAX5812 12-Bit Low-Power, 2-Wire, Serial Voltage-Output DAC Free Samples  
MAX6029 Ultra-Low-Power Precision Series Voltage Reference Free Samples  
MAX6033 Ultra-High-Precision SOT23 Series Voltage Reference  
MAX6173 High-Precision Voltage References with Temperature Sensor Free Samples  
MAX6381 SC70/µDFN, Single/Dual Low-Voltage, Low-Power µP Reset Circuits Free Samples  
MAX6495 72V, Overvoltage-Protection Switches/Limiter Controllers with an External MOSFET Free Samples  
MAX6720 Dual/Triple Ultra-Low-Voltage SOT23 µP Supervisory Circuits Free Samples  
MAX6746 µP Reset Circuits with Capacitor-Adjustable Reset/Watchdog Timeout Delay Free Samples  
MAX6950 Serially Interfaced, +2.7V to +5.5V, 5- and 8-Digit LED Display Drivers Free Samples  
MAX7032 Low-Cost, Crystal-Based, Programmable, ASK/FSK Transceiver with Fractional-N PLL Free Samples  
MAX7310 2-Wire-Interfaced 8-Bit I/O Port Expander with Reset Free Samples  
MAX8530 Dual Low-Dropout Linear Regulator with Reset or Low-Noise Output in UCSP or QFN  
MAX8569 200mA Step-Up Converters in 6-Pin SOT23 and TDFN Free Samples  
MAX8570 High-Efficiency LCD Boost with True Shutdown Free Samples  
MAX8667 1.5MHz Dual Step-Down DC-DC Converters with Dual LDOs and Individual Enables Free Samples  
MAX8808X 1A Linear Li+ Battery Chargers with Integrated Pass FET and Thermal Regulation in 2mm x 2mm TDFN Free Samples  
MAX8814 28V Linear Li+ Battery Charger with Smart Autoboot Assistant Free Samples  
MAX9618 High-Efficiency, 1.5MHz Op Amps with RRIO Free Samples  
MAX9718 Low-Cost, Mono/Stereo, 1.4W Differential Audio Power Amplifiers Free Samples  
MAX9850 Stereo Audio DAC with DirectDrive® Headphone Amplifier Free Samples  
MAX9939 SPI-Programmable Gain Amplifier (PGA) with Input VOS Trim and Output Op Amp Free Samples  
MAX9940 Signal-Line Overvoltage Protector for Low-Voltage Devices Free Samples  
MAX9945 38V, Low-Noise, MOS-Input, Low-Power Op Amp Free Samples  
MAX9945 38V, Low-Noise, MOS-Input, Low-Power Op Amp Free Samples  
MAXQ2000 Low-Power LCD Microcontroller  
MAXQ2010 16-Bit Mixed-Signal Microcontroller with LCD Interface  
MAXQ610 16-Bit Microcontroller with Infrared Module Free Samples  
MAXQ622 16-Bit Microcontrollers with Infrared Module and Optional USB Free Samples  
MAXQ8913 16-Bit, Mixed-Signal Microcontroller with Op Amps, ADC, and DACs for All-in-One Servo Loop Control  


Next Steps
EE-Mail Subscribe to EE-Mail and receive automatic notice of new documents in your areas of interest.
Download Download, PDF Format (216.5kB)  

© Jun 21, 2010, Maxim Integrated Products, Inc.
The content on this webpage is protected by copyright laws of the United States and of foreign countries. For requests to copy this content, contact us.

APP 4671: Jun 21, 2010
TUTORIAL 4671, AN4671, AN 4671, APP4671, Appnote4671, Appnote 4671